SOP for Verification of System Suitability test.

Standard Operating Procedure (SOP)
Here is a Standard Operating Procedure (SOP) for the Verification of System Suitability Test (SST). This procedure ensures that analytical systems perform according to predefined specifications before performing routine analysis in laboratories, particularly in pharmaceutical, clinical, and research laboratories.
1. Purpose
The purpose of this SOP is to establish the procedure for verifying the system suitability of analytical instruments before performing any routine analytical testing. This ensures that the equipment is functioning properly and that the results from subsequent analyses are reliable and accurate.
2. Scope
This procedure applies to the verification of system suitability tests for all analytical instruments used in routine testing in the laboratory. The SOP is applicable to chromatographic systems (e.g., HPLC, GC) and any other instruments that require periodic suitability verification.
3. Responsibilities
Laboratory Supervisor/Manager:
- Ensure that all personnel are trained and adhere to the system suitability test procedures.
- Review and approve system suitability results before analytical testing.
- Maintain records of all system suitability tests and verify the correctness of the results.
Analytical Chemists/Technicians:
- Perform system suitability tests according to this SOP.
- Document the results and notify the supervisor if the system fails to meet specifications.
Quality Control (QC) Personnel:
- Ensure that system suitability tests are conducted according to relevant guidelines.
- Review system suitability test results and approve the readiness of the system for analysis.
4. Definitions
System Suitability Test (SST): A test conducted to verify that the analytical system meets the required performance criteria before starting a specific analysis. This includes parameters such as precision, resolution, and sensitivity.
Instrument Calibration: The process of ensuring the instrument’s performance meets specific requirements through the use of known standards or reference materials.
Resolution: The ability of the system to distinguish between two closely eluting peaks or components in chromatographic analyses.
Precision: The degree of reproducibility or consistency in the results obtained from repeated measurements.
5. Procedure
5.1 Materials Required
- Standard Reference Materials or Calibration Standards
- System Suitability Test Solutions (e.g., test solutions or known standards)
- Chromatographic Instrument (e.g., HPLC, GC, etc.)
- Documentation Tools (e.g., system suitability log, data acquisition software)
- Pipettes, Vials, and Other Lab Supplies
5.2 System Suitability Test Preparation
Preparation of Test Solutions:
Prepare or procure suitable test solutions or standard reference materials with known concentration and composition according to the method’s specifications.
Ensure that all test solutions are properly labeled and stored under appropriate conditions.
Instrument Calibration:
Calibrate the analytical instrument (e.g., HPLC or GC) as per the manufacturer’s guidelines or the laboratory’s standard procedures.
Verify that all calibration standards are within their shelf life and have been stored properly.
Check Instrument Settings:
Ensure that the instrument settings, including flow rate, temperature, detection wavelength, and injection volume, are correctly configured as per the method’s requirements.
5.3 Performing System Suitability Test
Inject the System Suitability Solution:
- Inject the prepared system suitability test solution (typically a standard or test sample) into the chromatographic system.
- Make at least three injections to evaluate system performance.
Monitor Key Parameters: Evaluate the following system suitability parameters, as applicable to the analytical method:
Retention Time (RT): Verify that the retention time for the test sample or standard matches the expected values.
Peak Resolution: Ensure that the peaks in the chromatogram are well-resolved, with a resolution above the required threshold (typically ≥1.5).
Theoretical Plates (N): Check the number of theoretical plates to ensure the column is functioning efficiently (e.g., ≥2000 for HPLC).
Tailing Factor (T): Ensure that the tailing factor for the peaks is within acceptable limits (typically ≤ 2.0 for HPLC).
Area Consistency: Verify that the area of the peaks remains consistent across repeated injections, ensuring precision.
Signal-to-Noise Ratio: Verify that the signal-to-noise ratio is sufficient, typically ≥ 10:1 for low concentration analytes.
Document Results:
Record the results of the system suitability test in the appropriate format or system suitability log. Include relevant data such as retention times, peak areas, resolution, and other measured parameters.
Compare the measured results with the acceptance criteria defined in the method.
5.4 Acceptance Criteria
- Ensure the system suitability results meet the predefined acceptance criteria, which may include:
- Retention Time: ±2% deviation from the target
- Peak Resolution: ≥1.5 (for baseline separation)
- Theoretical Plates: ≥2000
- Tailing Factor: ≤2.0
- Signal-to-Noise Ratio: ≥10:1
- Precision: RSD (Relative Standard Deviation) ≤2%
5.5 In Case of Failure
Identify the Cause of Failure:
If the system suitability test fails to meet the criteria, investigate possible causes such as issues with the column, system components, or sample preparation.
Troubleshoot and Resolve:
Troubleshoot the instrument and resolve any identified issues (e.g., replace the column, adjust instrument parameters, or recalibrate the system).
Re-Test:
After resolving the issues, repeat the system suitability test to verify that the system is functioning correctly.
Document Failure:
If the system fails after troubleshooting, document the issue and report it to the laboratory supervisor or manager for further action.
5.6 Final Verification
System Ready for Use:
If the system suitability test passes, verify that the system is ready for routine sample analysis.
Document that the instrument is suitable for use.
Approval for Testing:
The laboratory supervisor or manager must review the system suitability test results and approve the system for routine use.
6. Documentation and Record Keeping
Maintain a System Suitability Log that includes:
- Date and time of the system suitability test
- Identification of the analytical system used
- Test solution details (concentration, batch number, etc.)
- Results for each key parameter (retention time, resolution, tailing factor, etc.)
- Any troubleshooting actions taken in case of failure
Signature and initials of the person performing the test and the supervisor/manager approving the results
Keep records for a minimum of [insert duration, e.g., 1 year].
7. Compliance and Auditing
Ensure that the system suitability test procedure adheres to relevant industry standards (e.g., USP, ICH, ISO) and regulatory guidelines.
The laboratory supervisor should regularly review system suitability results to verify ongoing compliance and to ensure the reliability of testing.
8. References
- USP Chapter <621> – Chromatography
- ICH Q2(R1) – Validation of Analytical Procedures
- Manufacturer’s User Manual for Instrument.
🎓 Discover one of the best Quality Assurance courses available — click below to explore the course that’s shaping future QA skills.