How is a halogen moisture analyzer calibrated?

  • Home
  • How is a halogen moisture analyzer calibrated?

How is a halogen moisture analyzer calibrated?

Calibrating a halogen moisture analyzer ensures its accuracy in measuring the moisture content of various substances. Calibration involves both temperature calibration and, optionally, weight (balance) calibration, depending on the model and application. Below is a detailed explanation of how this is typically performed:

Step-by-Step Calibration of a Halogen Moisture Analyzer

1. Preparation Before Calibration

  • Read the User Manual: Each model may have specific calibration procedures.

  • Allow Warm-Up Time: Turn on the device and let it warm up (usually 30 minutes) to stabilize the internal components.

  • Use a Stable Surface: Place the analyzer on a vibration-free, level surface.

  • Ensure Cleanliness: Clean the sample chamber, heating element, and weighing pan.

2. Weight Calibration (if required)

Many halogen moisture analyzers have built-in balances. You may need to calibrate the weighing mechanism:

Steps:

  • Access the balance calibration menu.

  • Place a certified calibration weight (typically 50g or 100g depending on the model).

  • Follow the on-screen prompts to complete the calibration.

Note: Use only certified Class F1/F2 calibration weights for accuracy.

3. Temperature Calibration

Temperature calibration is essential because the moisture analyzer uses halogen lamps to heat the sample, and inaccurate temperature readings can affect results.

Materials Needed:

  • A temperature calibration kit (usually includes a temperature sensor and test material or pad).

  • Some analyzers have built-in routines; others require external tools.

Procedure:

  1. Enter the temperature calibration mode via the settings menu.

  2. Place the temperature sensor or calibration standard in the sample area.

  3. The device heats to a preset target temperature (e.g., 100°C, 160°C).

  4. The analyzer compares actual vs. expected temperatures.

  5. It then adjusts its internal measurement system accordingly.

Important: If using an external calibration kit, the sensor must be traceable and accurate to at least ±1°C.

4. Moisture Method Verification (Optional but Recommended)

This is not calibration per se, but a check:

  • Use a known standard sample (e.g., sodium tartrate dihydrate with 15.66% moisture).

  • Run the test to see if the result matches the known value.

  • If deviations are consistent, you may need to adjust method parameters or recalibrate.

5. Documentation

Always document:

  • Date/time of calibration

  • Calibration weights used (ID and class)

  • Temperature calibration results

  • Any deviations or corrective actions

 Tips for Accurate Calibration

  • Calibrate regularly, especially before critical measurements.

  • Avoid drafts and vibrations during calibration.

  • Perform temperature calibration monthly or as recommended.

  • Keep a calibration log for audits and quality assurance.