Calibration of Volumetric Glassware used in Pharmaceuticals

  • Home
  • Calibration of Volumetric Glassware used in Pharmaceuticals

Calibration of Volumetric Glassware used in Pharmaceuticals

Calibration of Volumetric Glassware in Pharmaceuticals

1. Introduction

Volumetric glassware (such as pipettes, burettes, volumetric flasks, measuring cylinders) is widely used in pharmaceutical laboratories for accurate measurement of liquid volumes in analytical testing, solution preparation, and standardization procedures.
Calibration ensures these glassware items deliver or contain the correct volume within specified tolerance limits, as per pharmacopeial and ISO guidelines.

2. Objective

  • To verify the accuracy of volumetric glassware against traceable standards.

  • To ensure measurement reliability in analytical work, minimizing systematic errors.

  • To comply with GMP, GLP, and regulatory requirements.

3. Principle

The calibration of volumetric glassware is based on gravimetric analysis:

  • A known volume of distilled or deionized water is measured using the glassware.

  • The water is weighed using an analytical balance.

  • Using the density of water at a given temperature, the actual volume is calculated.

  • The measured value is compared to the nominal (marked) value to determine the error.

Key formula:

Volume (mL)=Weight of water (g)Density of water at measured temperature (g/mL)\text{Volume (mL)} = \frac{\text{Weight of water (g)}}{\text{Density of water at measured temperature (g/mL)}}

4. Equipment & Materials

  • Analytical balance (calibrated)

  • Distilled or deionized water

  • Thermometer (for water temperature measurement)

  • Clean, dry volumetric glassware

  • Density table of water (as per ASTM or USP)

5. Procedure

A. General Steps

  1. Clean & Dry the glassware: Remove grease or dust using a suitable cleaning solution, then rinse with distilled water.

  2. Weigh Empty Vessel: Record the weight (W₁).

  3. Fill to Mark: Fill with distilled water at ambient laboratory temperature (close to 20°C or as per requirement), ensuring the meniscus aligns exactly with the calibration mark.

  4. Weigh Filled Vessel: Record the weight (W₂).

6. Tolerance Limits (Example: ISO 4787 / USP)

Glassware Tolerance (± mL) for 100 mL
Volumetric Flask 0.10 mL
Pipette (Class A) 0.08 mL
Burette (Class A) 0.05 mL
Measuring Cylinder (Class A) 0.50 mL

7. Calibration Frequency

  • New Glassware: Before first use.

  • In-use Glassware: At least annually or as per SOP.

  • After repair or suspected damage.

8. Documentation

  • Glassware ID/Serial number

  • Date of calibration

  • Nominal capacity

  • Temperature of water

  • Actual volume measured

  • % error and pass/fail status

  • Signature of analyst and reviewer

9. Regulatory References

  • USP <31>, <41>, and <1251>

  • ISO 4787: Laboratory Glassware – Volumetric Glassware – Methods for Use and Testing

  • ICH Q7 & Q10 (Equipment Calibration)

  • WHO Good Laboratory Practices