Calibration of Brookfield Viscometer

  • Home
  • Calibration of Brookfield Viscometer

Calibration of Brookfield Viscometer

Calibration of Brookfield Viscometer in Pharmaceuticals

1. Introduction

A Brookfield viscometer is widely used in pharmaceutical industries to measure the viscosity of semi-solids, creams, gels, and certain liquid formulations.
Regular calibration ensures accuracy, reproducibility, and compliance with GMP and pharmacopoeial requirements.

2. Objective

  • To verify the accuracy of viscosity readings.

  • To ensure instrument performance is within acceptable limits.

  • To maintain traceability to certified standards.

3. Principle

Calibration involves comparing viscosity readings of certified viscosity standard fluids (traceable to NIST) with their known values at a controlled temperature.
If the observed values fall within the acceptable tolerance range specified by the manufacturer (usually ±1% for Brookfield), the viscometer is considered calibrated.

4. Equipment & Materials

  • Brookfield viscometer (LV, RV, or appropriate model)

  • Certified viscosity standard fluids (e.g., Brookfield Silicone Oil Standards)

  • Temperature-controlled water bath

  • Appropriate spindles for the instrument model

  • Stopwatch (if needed for manual models)

5. Procedure

  1. Preparation

    • Clean the spindle and sample chamber with a suitable solvent and dry completely.

    • Set the viscometer to the correct speed and spindle as per manufacturer’s instructions.

    • Maintain the test fluid at the specified calibration temperature (commonly 25 ± 0.1 °C).

  2. Measurement

    • Immerse the spindle into the viscosity standard fluid up to the immersion groove.

    • Start the viscometer and allow readings to stabilize (as per instrument guidelines).

    • Record the viscosity reading.

  3. Comparison

    • Compare the measured viscosity to the certified value of the standard at the same temperature.

    • Repeat for all required viscosity ranges using different standards.

  4. Acceptance Criteria

    • Reading should be within ±1% of full-scale range (or as specified by the manufacturer).

    • If results are outside the limit, perform maintenance/servicing and recalibrate.

6. Calibration Frequency

  • Before first use

  • Annually (or as per SOP)

  • After repairs, spindle replacement, or suspected malfunction

7. Documentation

  • Viscometer ID and model

  • Calibration date

  • Standard fluid lot number and certified value

  • Spindle number, speed, and temperature

  • Observed viscosity and % deviation

  • Pass/Fail status

  • Signatures of analyst and reviewer

8. Regulatory References

  • Brookfield Viscometer Operating Manual

  • USP <912> Viscosity

  • ISO 2555: Plastics — Resins in the Liquid State or as Emulsions or Dispersions — Determination of Apparent Viscosity Using a Rotational Viscometer

Submit a Comment

Your email address will not be published. Required fields are marked *


The reCAPTCHA verification period has expired. Please reload the page.

Hello
Chat now via Whatsapp