Extraneous Peaks in Chromatographic Analysis

Extraneous Peaks in Chromatographic Analysis
In chromatographic techniques like HPLC and GC, extraneous peaks are unwanted peaks in the chromatogram that are not related to the analyte(s) of interest, mobile phase, or expected impurities. They can distort quantitative and qualitative results, cause system suitability failures, and trigger investigations in pharmaceutical quality control.
Causes of Extraneous Peaks
1. Instrument-Related Causes
-
Carryover from previous runs
-
Incomplete flushing of sample loop, injector, or column.
-
-
Contamination in injector or autosampler
-
Residual analytes or precipitated samples.
-
-
Detector noise or instability
-
Lamp deterioration (UV detector), thermal noise (RID, FLD).
-
-
Leaching of materials
-
Tubing, seals, or septa leaching compounds into the mobile phase.
-
-
Ghost peaks from column memory
-
Retained compounds eluting in subsequent runs.
-
2. Mobile Phase & Solvent Causes
-
Impurities in solvents
-
Non-HPLC grade solvents containing UV-absorbing compounds.
-
-
Mobile phase contamination
-
Improper storage, microbial growth (especially in aqueous buffers).
-
-
Air bubbles in system
-
Baseline disturbance leading to spike-like peaks.
-
-
Degradation products
-
Mobile phase components breaking down under light/heat.
-
3. Sample-Related Causes
-
Sample matrix interference
-
Excipients or formulation components eluting unexpectedly.
-
-
Decomposition of analyte
-
Breakdown into secondary peaks during analysis.
-
-
Improper filtration
-
Filter extractables appearing as peaks.
-
-
Residual diluent contamination
-
Volatile residues introducing foreign compounds.
-
4. Environmental Causes
-
Laboratory air contamination
-
Volatile organics from solvents, perfumes, cleaning agents.
-
-
Dust or particulate matter
-
Entering mobile phase or sample vials.
-
-
Temperature and humidity changes
-
Affecting baseline stability in GC.
-
Identification of Extraneous Peaks
-
Run blank injections
-
Mobile phase blank and diluent blank to check for contamination.
-
-
System flushing
-
High-strength solvent flushing to see if peaks disappear.
-
-
Column changeover
-
Replacing with a new or different column to rule out column-related carryover.
-
-
Peak purity analysis
-
Using diode-array detection (DAD/PDA) or mass spectrometry.
-
Prevention and Control
-
Use high-purity solvents and reagents (HPLC/GC grade).
-
Regularly clean and maintain injector, sample loop, column, and detector.
-
Filter mobile phase through 0.45 µm or 0.22 µm filters.
-
Avoid mobile phase degradation (protect from light, prepare fresh buffer).
-
Use dedicated glassware for chromatographic analysis.
-
Run system suitability and blank checks before analysis.
-
Implement column wash protocols after high-retention samples.
Regulatory Consideration
In pharmaceutical QC, extraneous peaks must be investigated under data integrity and OOS guidelines. If a peak appears in blank injections or replicates, root cause must be determined before release decisions.